42
3.2. Maintenance
Window Cleaning (LISST-Infinite)
Step 1
Disconnect the turbid water supply, so that water will not flow through
the system while you clean the optics. Disconnect the power so that the
instrument does not attempt to sample while you clean. Then open the
front door on the LISST-Infinite.
Step 2
Drain the water out of the system by pressing in solenoid valve #3.
Step 3
Loosen the screws holding the optics cell port in place
Step 4
Remove the port. Be careful not to lose the screws!