Click the touch screen, then calibrating touchscreen is
finished.
13. Common Faults and Treatment
Below are the possible faults of instrument and solutions:
Table 8-1 Fault and Treatment
Faults
Causes
Solutions
Instrument fails to
start
Out of battery
Charge the battery
Not connected to
external
power supply
Use AC/DC Adapter to
supply power.
Button fails to
respond
Check if any
button is kept
pressed.
Make sure no button
is kept pressed.
Length of optical
fiber measured is not
accurate
a. Refractive
index of optical
fiber is not set
accurately
b. Fiber cable
’
s
coefficient of
correction is not
set
accurately
a. Set the refractive
index of optical fiber to
the refractive index
calibrated by fiber
manufacturer.
b. Set fiber cable
’
s
coefficient of
correction to
1.0000 or to the
coefficient of
correction
calibrated by fiber
cable manufacturer.