SIE-CNTEP-1100002104-2014
LCFS
AV6418 OTDR
58
optical fiber under test).
l
IOR:
The IOR of optical fiber and optical cable can be obtained
from production factory. The value of IOR is set to:
1.00000~2.00000.
Inaccurate IOR setup will result in inaccurate optical fiber
length.
l
Optical Cable Correction:
The setup of optical cable correction coefficient depends on
the errors between optical fiber length and optical cable
length after optical fiber is made into cable. It can be
obtained from optical cable production factory. It can be set
to: 0.80000-1.00000. The default value is 1.00000.
Inaccurate optical cable correction setup will result in
inaccurate optical cable length.
l
Length Unit:
The instrument allows users to set length unit to km or kft.
The default length unit is km.
l
Auto Analyse after Sampling:
When this item is selected, in real-time, average and dead
zone test mode, the instrument will analyze trace, mark
Página No.70 LCFS
CONVALIDACIÓN DE ERRORES DEL PROCESO:
“ADQUISICIÓN DE OTDR (OPTICAL TIME DOMAIN REFLECTOMETER) – REDES GPON/
FTTH/ SDH"