AV6418 OTDR
76
change and set the mode option. Finally press
key to
confirm the setting and exit from the setup interface back to
the main screen; press
key to cancel the setting back
to the main screen.
(1) When other test conditions are the same, the value of
attenuation is set as 0dB, the testable optical fiber distance
is longest, the signal to noise ratio of trace is relative good,
but it may make the OTDR near-end trace saturated so that
the near-end trace can not be observed.
(2) When other test conditions are the same, the value of
attenuation is set as 20dB, the testable optical fiber distance
is longest, and the signal to noise ratio of trace is poor, but it
can achieve smaller test blind zone.
(3) Under the manual test mode (that means real-time test
mode and average test mode), the test attenuation should
be selected according to the optical fiber to be tested.
(4) When
“
Track history test
”
Mode is selected, and the test
range is not less than 64km (that means when measuring
relative
long
optical
fiber
link),
the
instrument
will
automatically adjust attenuation setup and display the
complete optical fiber link near-end and far-end trace on the
display.