SIE-CNTEP-1100002104-2014
LCFS
AV6418 OTDR
120
Ø
Expand the wave form in the horizontal direction and
set the mark point A at the end of the linear zone which is
ahead of the tested reflection point.
Ø
Locate the mark point B at the top of the reflection
event point, as shown in Figure 8-34.
Ø
At this moment the reflection loss value displayed on
the operating window is the reflection loss of the tested
reflection point.
Figure 8-34 Reflection loss test
Distance (A-B):
0.03440 km:
The distance between cursor
A and cursor B
Reflection loss:
42.472 dB:
The reflection loss of reflection
peak of cursor A and cursor B
Página No.132 LCFS
CONVALIDACIÓN DE ERRORES DEL PROCESO:
“ADQUISICIÓN DE OTDR (OPTICAL TIME DOMAIN REFLECTOMETER) – REDES GPON/