SIE-CNTEP-1100002104-2014
LCFS
AV6418 OTDR
8
Use of OTDR
Never connect an optical fiber carrying any kind of optical
signal to the test port of AV6418 OTDR. Otherwise, it could
result
in
inaccurate
measurement
result
and
even
permanent damage to the instrument. Please make sure
that all the optical fibers under test are inactive.
Precautions for Use of Instrument
AC/DC
Adapter
Please use the adapter provided by Shandong Senter
Electronic Co., Ltd for power supply. Use of other type of
adapter can damage the instrument.
Internal
Battery
(1) For the best battery performance, the instrument shall be
powered with internal battery first at the beginning of use
until the battery is exhausted, and then it can be charged
2~3 times.
(2) If the battery has been in storage for a long time, it shall
be charged and discharged circularly as above for the first
time use to achieve an optimum performance.
(3) In order to prolong the life of internal battery, remove
external power supply after charging. For a long-time use of
external power supply, please remove the internal battery
and keep it properly, before that, please charge it first, and
make charge-discharge cycle once every 3 months to
Página No.20 LCFS
CONVALIDACIÓN DE ERRORES DEL PROCESO:
“ADQUISICIÓN DE OTDR (OPTICAL TIME DOMAIN REFLECTOMETER) – REDES GPON/
FTTH/ SDH"