12.5 Sensor Head Backwards Compatibility
The Conquest 100 is fully compatible with earlier versions of the Conquest Sensor Head and
cable so if you have a Conquest Enhanced or Conquest SL Sensor Head available, try
swapping these parts in one at a time and see if you can isolate the problem or solve it.
12.6 Warning Beeps
When acquiring data (either in Line Scan or Grid Scan modes), if the system starts to
“beep”
when it’s not supposed to, there are two possible causes:
1. The operator is holding down the
Enter
button or Display Unit key too long when starting
the Line Scan. The system may register this as two button presses, resulting in the
system starting and stopping immediately. In this instance the triple beep is occurring
because the scan is being stopped. A lighter touch on the button is necessary.
2. The Sensor Head is being pushed too fast. In this case, the DynaQ will also show yellow
and white. To ensure good data quality, simply slow down the speed of the Sensor
Head. In Grid Scan mode, you may be prompted to re-collect that line.
12.7 Sensor Head Keypad Doesn't Respond
If the Sensor Head keypad does not respond, the usual reason is that the Sensor Head has
come unplugged during operation. Power down the system, reconnect the Sensor Head and
restart to restore full keypad functionality.
12.8 Constant Prompt to Perform System Test
If the operator is constantly prompted to run the System Test for the Sensor (
resulting in the inability to collect data, turn off IEP (
5.1.10) in the Preferences menu. If
IEP is disabled, allow the system to warm up for 5 minutes before starting data collection. Also
note that the next time Conquest is powered on, IEP defaults to ON.
If the system is constantly prompting to perform a system test, it may indicate a problem with
the equipment. Scanning can be continued with IEP turned off but if the problem persists, it is
recommended that Sensors & Software Inc. be contacted to help remedy the problem.
12.9 Collecting Test Data
One of the best ways of detecting problems with the Conquest system is by comparing data with
data collected previously.
Soon after receiving the system and getting comfortable with its operation, collect a
24”x24”
(600x600mm) grid of data at a convenient, easily accessible location. The grid should be saved
electronically and perhaps plotted out on paper and dated. The test grid could be collected say,
Содержание CONQUEST 100
Страница 1: ... Copyright 2015 Sensors Software Inc 2015 00049 05 ...
Страница 2: ......
Страница 4: ......
Страница 10: ......
Страница 16: ...Conquest 100 Overview 6 ...
Страница 46: ...Conquest 100 Main Screen 36 ...
Страница 54: ...Conquest 100 Tools Setup 44 On the page that follows ensure the setting Allow less secure apps is set to ON ...
Страница 82: ...Conquest 100 Line Scan 72 Figure 6 16 Adding Interps over rebar Figure 6 17 Available colors for Interps ...
Страница 84: ...Conquest 100 Line Scan 74 ...
Страница 104: ...Conquest 100 SliceView 94 ...
Страница 114: ...Conquest 100 Transferring Data to a PC 104 ...
Страница 116: ...Conquest 100 Transferring Data to a PC 106 ...
Страница 120: ...Conquest 100 Troubleshooting 110 ...
Страница 124: ...Conquest 100 Care and Maintenance 114 ...
Страница 126: ...Conquest 100 Parts List Accessories 116 14 2 Accessories The following accessories are available for purchase ...
Страница 128: ...Conquest 100 Technical Specifications 118 ...
Страница 130: ...Conquest 100 Appendix A Health Safety Certification A2 ...
Страница 140: ...Conquest 100 Appendix C Instrument Interference C2 ...
Страница 142: ...Conquest 100 Appendix D Safety Around Explosive Devices D1 ...