MEDICAL DIVISION

This document is property of Seiler Instrument & Mfg. Co., Inc.  

No part of this manual should be reproduced or transmitted without the expressed written consent being 

obtained. All the figures and images appearing in this manual are for illustration purposes only and 

may vary according to the version of the device purchased. The information contained in this manual is 

subject to change without notice. Please contact Seiler whenever additional information is needed.  

This device is restricted to the sale and the use by authorized and trained personnel.

COLPOSCOPE 935

USER’S MANUAL

Including installation instructions for all mounting 

options

Содержание COLPOSCOPE 935

Страница 1: ...this manual are for illustration purposes only and may vary according to the version of the device purchased The information contained in this manual is subject to change without notice Please contac...

Страница 2: ...n 12 2 0 Symbols Used in this Manual or on the Device 13 3 0 Hazards Associated with Improper Installation 14 3 1 Installation Instructions Preparing and Siting 15 3 2 Installation Instructions 16 4 0...

Страница 3: ...3 SYMBOLS USED IN THIS MANUAL OR DEVICE 25 26 WARNINGS 27 WARRANTIES 28 29 CERTIFICATE OF CONFORMITY 30 DECLARATION OF CONFORMITY 31...

Страница 4: ...d ISO standards We also ensure all optical components have undergone an anti reflective multi layer treatment guaranteeing the efficiency of the system and enhancing the longevity of all of our produc...

Страница 5: ...details of its operation prior to use Therefore it is advisable that personnel operating the colposcope read this manual prior to using the equipment colposcope floor stand NOMENCLATURE OF THE NUMBERE...

Страница 6: ...f the objective is too blotched change to another cotton swab for each circular movement to avoid spreading the impurities again The Objective lens can be protected by use of the Sterilizable Protecto...

Страница 7: ...only one image PARAFOCALIZING PROCEDURE The binocular eyepieces offer a diopter adjustment to allow use of the colposcope either with or without corrective lenses However the diopter adjustment can on...

Страница 8: ...ire colposcope While looking through the eyepieces move the colposcope in a direction farther away then closer to the area of interest Note the direction that improves focus then continue moving the c...

Страница 9: ...71 32 38 9 33 7 41 22 50 28 33 10 67 8 47 18 00 22 67 13 33 10 59 15 00 18 89 16 00 12 71 11 25 14 17 21 33 16 94 400 136 OBJECTIVE LENS 16x 16 20x 12 illuminated field 59 magnification field 10x 18m...

Страница 10: ...ce complies with International Standards as set forth in manufacturer s declaration following for electromagnetic compatibility for medical electrical equipment and or systems These standards are desi...

Страница 11: ...surgical microscope for instance Operation medical or dental training in surgical procedures dental health procedures plus specific training by the supplier of the larger medical system surgical micro...

Страница 12: ...iated Immunity 3 V M 8o Mhz to 2 7 Ghz in 1 increments professional use 80 Sinusoidal AM at 1000 Hz Pass A1 2007 Radiated Immunity Pass A2 2010 Radiated Immunity Pass EN 61000 4 4 2012 EFT Pass IEC 61...

Страница 13: ...s important information regarding set up and operation to facilitate ease of use and obtain effective results WARNING This term contains critical information by identifying conditions or practices tha...

Страница 14: ...ion distance between the unit and any device which may be affected by the electromagnetic energy coming from the unit Power the unit from a circuit different from any circuit that contains a device th...

Страница 15: ...ed Should it be used it is the installer s responsibility to ensure it does not block airflow AND is not a flammable material Failure to follow this instruction could result in a fire Do not operate t...

Страница 16: ...on Consistent noise no scraping clicking etc No interference with nearby critical equipment No interference from operation of nearby equipment large unshielded motors etc Power down installation is co...

Страница 17: ...the MAINS ON OFF blue switch with international 0 1 designation The blue light should be off Then remain the MAINS cable Consult the factory with information as to why the installation was halted Do...

Страница 18: ...shall inspect the unit for internal damage Employ only authorized and properly trained personnel to perform maintenance functions 5 1 FIELD MAINTENANCE INSTRUCTIONS READ FULLY BEFORE ATTEMPTING FIELD...

Страница 19: ...ming level by pressing UP or DOWN buttons Shift to RUN mode by pushing the RUN STANDBY mode button Confirm functionality Free flow of air Illumination Consistent noise no scraping clicking etc No inte...

Страница 20: ...e once it has been determined to be at its end of life should be done in accordance with the fol lowing Ensure it is not connected to MAINS power Clean thoroughly with anti bacterial solution to mitig...

Страница 21: ...21 8 2 FRONT PANEL CAMERA POWER Jack 12 VDC at 500 ma max For installation of LIGHT CABLE 8 3 rear panel IFC Mains Power Inlet Mains ON OFF Switch Exhaust Fan Factory Connector Do Not Use...

Страница 22: ...Release this jumps intensity b Press and Hold this will cause the device to slowly scroll through all available dim ming levels The top line of the display identifies the company or brand The bottom l...

Страница 23: ...nt part with out charge Should your LED Light Source product s need servicing under this warranty please contact Seiler Instru ment return authorization documentation You should carefully pack unit in...

Страница 24: ...________________________________________ ______________________________________________________________________________________ ________________________________________________________________________...

Страница 25: ...ccompanying documents or Attention see instructions for use Symbol for Caution Hot Surface This symbol cautions that surface should not be touched as it may be hot and could be a burn hazard Symbol ca...

Страница 26: ...ean Union EU States this symbol should be used to mark devices that are reusable and not contaminated at the end of the device life This symbol is a mandatory marking for devices entering the European...

Страница 27: ...condition which could cause Injury to personnel or damage to equipment Do not tilt the colposcope Use the colposcope on flat surfaces only Horizontal and vertical forces create instability and may re...

Страница 28: ...instruction for the use of the product abuse physical mishandling or natural causes such as flood fire earthquake or other perils as determined by SIC or 2 if any repairs are made by persons unauthori...

Страница 29: ...n for the use of the product abuse physical mishandling or natural causes such as flood fire earthquake or other perils as determined by SIC or 2 if any repairs are made by persons unauthorized by SIC...

Страница 30: ...30...

Страница 31: ...31...

Страница 32: ...Louis Missouri 63122 USA Toll Free 800 489 2282 Local 314 968 2282 Email micro seilerinst com www seilerinst com Atlantico Systems Ltd 34 Oldfield Kingston Galway Ireland www atlanticosystems com MED...