8.4 controls
CONTROL PAD
LCD DISPLAY
There are (3) basic controls.
An UP button for increasing intensity.
A DOWN button for decreasing intensity.
A LAMP button for toggling the device between
STANDBY mode and RUN mode. Standby
mode will remove power to the LED / Extinguish
the illumination but will leave other circuits active
and is intended to provide a faster response time
than the MAINS ON/OFF switch (between proce-
dures).
The UP and DOWN buttons can be operated in
two modes:
a) Press and Release – this “jumps”
intensity.
b) Press and Hold – this will cause the
device to slowly scroll through all available dim-
ming levels.
The top line of the display identifies the company
or brand.
The bottom line displays status as well as dim-
ming level.
The bottom line will indicate STANDBY or RUN.
Each SQUARE represents 10% of the range (ie
3 SQUARES ~ 30% of maximum illumination).
Should the device detect a fault it will indicate
FAULT followed by the type of fault.
8.5 Pole mounting
POLE CLAMP
There are (4) threaded holes located underneath
the illuminator. Caution must be exercised to use
screws that project into the chassis NO MORE
THAN ⅛”. Exceeding this can cause damage to
the internal circuitry.