When to clean and disinfect the analyzer
We recommend that you clean and disinfect the analyzer once a week. The cleaning and
disinfection instructions were validated for 10,950 cleaning and disinfection cycles (10,950
cleaning wipes and 10,950 disinfecting wipes) on the analyzer which supports cleaning and
disinfection after each patient, up to 10 patients per day, for the 3 year use life of the analyzer.
Please examine the LCD, test strip port, buttons, and surface of your analyzer after the cleaning
and disinfecting cycles. Stop using the analyzer and contact our customer service (1-855-486-
2888) immediately if you find the following:
•
Thin silver streaks appear, become cracked, soft, dissolved, brittle, or swollen; clouding of
LCD display
•
You are unable to turn on/off your meter or operate left/right button
•
You are unable to enter the meter settings or function modes, or you are unable to recall
your testing results
Cleaning STEPS
STEP 1- Dispose of used test strip.
STEP 2 - Take a wipe from the package or
container. If the wipe is very wet, wring
it slightly to discard excess liquid.
STEP 3 - Wipe each side of your analyser, according to the following method:
<Appearance of analyser> <Inside and under the door>
1)
Wipe the outside of the analyzer up and down 4 times.
2)
Wipe the outside of the analyzer left and right 4 times.
3)
Open the door, and then repeat steps 1 and 2 for the inside and under the
door.