Before you start a test
CAUTION
Carefully read and follow the instructions and package inserts for the lipid/glucose test strips
and control solutions. It is very important to follow the instructions in order to prevent an
incorrect result or improper treatment.
Samples
For measuring lipid (cholesterol)
The SD LipidoCare Professional Analyzer for lipid testing is designed for testing a fresh capillary
whole blood sample from the fingertip.
For measuring glucose
The SD LipidoCare Professional Analyzer for glucose is designed for testing a fresh capillary
whole blood sample from the fingertip, palm, upper arm, or forearm.
Refer to the Safety Information
There is a potential risk of infection. To clean and disinfect this analyzer, please refer to the
Chapter 6 Maintenance and Troubleshooting section.
Operating conditions
To ensure proper function of the SD LipidoCare Analyzer, observe the following guidelines:
Operate the analyzer within the acceptable temperature and the humidity range. This
range is test-dependent:
For Lipid Testing : 64–90°F (18–32°C), 10–90%RH
For Glucose Testing: 50–113°F (10–45°C), 10–90%RH
To perform a measurement, place the analyzer on a level surface or hold it in your
hand.