STEP 4 - Allow the analyzer to air dry. Dispose of the wipe.
Disinfecting STEP
STEP 1
Take another wipe and then wipe the analyzer thoroughly.
The analyzer should be cleaned prior to the disinfection step.
•
Wipe the outside of the analyzer up and down 4 times.
•
Wipe the outside of the analyzer left and right 4 times.
•
Open the door and then repeat steps 1 and 2 for the inside and under the
door.
STEP 2
Allow the surface to remain wet for 1 minute.
STEP 3
Allow to air dry.
STEP 4
After handling the analyzer or test strips for cleaning and disinfection, wash
your hands thoroughly with soap and water.
CAUTION
Do not get water inside the analyzer. Never immerse the analyzer or hold it under
running water, because it will damage the analyzer.
Do not spray directly into the strip slot! Moisture in the strip slot may lead to incorrect
results. If you suspect that moisture may have entered the strip slot, perform a lipid or
glucose control solution test.
Do not wrap the meter in a wipe; take extreme care not to get liquid in the test strip slot
or cable port of the analyzer.
After disinfection, users’ gloves should be removed and hands should be thoroughly
washed with soap and water before proceeding to the next patient.
If you have any questions, please contact our customer service center at 1-855-486-2888.