2.
Procedure for the lipid control solution
1)
Prepare the lipid test strip and the code chip.
2)
Make sure the code chip matches the code number of the test strip package.
3)
Insert a test strip into the analyzer.
4)
When the OPE message is displayed on the screen, open the door. Press the left button and
hold it for 3 seconds, then check to ensure that the control solution icon appears on the
screen. If you don't want to do a control solution test, press the left or right button one time
to exit.
5)
Gently shake the lipid control solution and apply a drop to the test strip, using a capillary
tube.
6)
After 3 minutes, the test result will appear on the screen.
7)
Compare the result to the range printed on the control solution insert. If the results are not
within the lipid control range printed on the insert, the analyzer and the test strip may not
be working properly. Repeat the control test. If the control solution result is still not
acceptable, please contact our customer service center at 1-855-486-2888.
8)
Remove the used test strip from the analyzer and discard it.
CAUTION:
•
Store upright and refrigerate SD Lipid Control Solution at 36–46˚F. It can be expected to
give stable results through the expiration date printed on the label. Minimize exposure to
strong light.
•
The SD Lipid Control Solution should be tested at room temperature, 64–90°F (10–90%
RH).
•
Do not use the SD Lipid Control Solution if it is
cloudy
has an odor
shipped or stored improperly
•
The acceptable range printed on the lipid control solution insert is for the SD Lipid
Control Solution only. It is not a recommended range for your lipid profile levels.