CHAPTER 10. TECHNICAL INFORMATION
1.
Analyzer Specifications
Battery Operation
4 AAA 1.5 V, Alkaline batteries
Battery Life
Around 1,000 tests
Display
LCD
Controls
4 Buttons (SET/PRT, ON/OFF, arrow: >/<)
Memory
500 results
Automatic shut-off
•
1 minute after last user action without a test strip inserted into
the analyzer
•
5 minutes after last user action with a test strip inserted into
the analyzer
2.
Test Strips
Measuring Range
•
TC: 100–450 mg/dL ( 2.59–11.64 mmol/L )
•
TG: 45–650 mg/dL ( 0.51–7.34 mmol/L )
•
HDL: 25–95 mg/dL ( 0.65–2.46 mmol/L )
•
Glucose: 20–600 mg/dL (1.2–33.3 mmol/L)
Sample
Lipid
Fresh capillary whole blood from the fingertip,
Venous whole blood, or Serum or plasma
Glucose
Fresh capillary whole blood from the fingertip,
palm, forearm, or upper arm
Sample Size
Lipid
35uL
Glucose
0.9uL
Test Time
Lipid
3 minutes
Glucose
5 seconds
Storage Temperature
36–90°F (10–90% RH) for both lipid and glucose test strip
Hematocrit
•
TC, TG: 30–55%
•
HDL: 30–52%
•
Glucose: 20–60%
Operating Condition
Lipid
64–90°F(10–90% RH)
Glucose
50–113°F(10–90% RH)
Altitude
Up to 3,776m (12,388 ft) only for glucose
- 54 -