60 | 8 4
--
Example:
Set the upper current for DCW test in STEP1 to 1mA
Command message: FUNC:SOUR:STEP 1:DC:UPPC 1
Query message: FUNC:SOUR:STEP 1:DC:UPPC?
Return value: 1.000
FUNC:SOURce:STEP:DC:LOWC
To set /inquiry about the LOWER current for DCW test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:DC:LOWC<current>
Queery message:FUNC:SOUR:STEP <sn>:DC:LOWC?
--
Data<current>:
Data format: float
Data range: 0
~
upper current (0 is OFF)
Data accuracy: 0.0001mA
Data unit: mA
--
Example
:
Set the lower current for DCW test in STEP1 to 1mA
Command message: FUNC:SOUR:STEP 1:DC:LOWC 1
Query message: FUNC:SOUR:STEP 1:DC:LOWC?
Return value: 1.000
FUNC:SOURce:STEP:DC:TTIM
To set /inquiry about the TEST time for DCW test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:DC:TTIM<time>
Query message: FUNC:SOUR:STEP <sn>:DC:TTIM?
--
Data<time>:
--Data format: float
Data range: 0, 0.3
~
999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s
--
Example:
Set the test time for DCW test in STEP1 to 1s
Command message: FUNC:SOUR:STEP 1:DC:TTIM 1
Query message: FUNC:SOUR:STEP 1:DC:TTIM?
Return value: 1.0
FUNC:SOURce:STEP:DC:RTIM
To set /inquiry about the RISE time for DCW test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:DC:RTIM<time >
Query message: FUNC:SOUR:STEP <sn>:DC:RTIM?
--
Data<time>:
Data format: float
Data range: 0~999.0s (0 is OFF)
Data accuracy: 0.1s
Data unit: s