Table 4-3
4.3.2 DC Withstanding Voltage Test Mode
Figure 4-13 DC Withstanding Voltage Setup
Setting item
Range
Default
Explanation
MeasVolt
0.050
~
6.000kV,
0 — ClOSE
CLOSE
Set the voltage required for DC withstand
voltage testing.
Hi Limit
0.0001mA
~
25mA
500.0uA
Set the high limit of the test current. The
voltage is < 1.5kV, the maximum current
can reach 20mA; the voltage is
≥
1.5kV, the
maximum current can reach 25mA.
Low Limit
0.0001mA
~
Upper
Limit,
0 —CLOSE
CLOSE
Set the low limit of the test current. The low
limit must be
≤
the high limit or Close.
Arc Limit
1mA
~
10mA,
0 —CLOSE
CLOSE
Set the allowable DC arc current high limit.
Close means no requirement for arc.
Arc Ramp
1mA
~
10mA,
0 —CLOSE
CLOSE
Set the high limit of the arc current when
the DC voltage rises.
Close means no requirement for arc when
voltage rises.