5.2.3.5 Commands for IR Setup Function
FUNC:SOURce:STEP:IR:VOLT
To set /inquiry about the voltage for IR test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:IR:VOLT<voltage >
Query message: FUNC:SOUR:STEP <sn>:IR:VOLT?
--
Data<voltage>:
Data format: float
Data range: 50~1000
Data accuracy: 1
Data unit: V
--
Example:
Set the voltage for IR test in STEP1 to 1000V.
Command message: FUNC:SOUR:STEP 1:IR:VOLT 1000
Query message: FUNC:SOUR:STEP 1:IR:VOLT?
Return value: 1000
FUNC:SOURce:STEP:IR:UPPC
To set /inquiry about the UPPER resistance for IR test.
--
Syntax
:
Command message FUNC:SOUR:STEP<sn>:IR:UPPR <resistance >
Query message: FUNC:SOUR:STEP <sn>:IR:UPPR?
--
Data< resistance>:
Data format: float
Data range: 0, lower limit~50.0G
Ω
(0 is OFF)
Data accuracy: 1.0 M
Ω
Data unit: M
Ω
--
Example:
Set the upper resistance for IR test in STEP1 to 1 M
Ω
.
Command message: FUNC:SOUR:STEP 1:IR:UPPR 1
Query message: FUNC:SOUR:STEP 1:IR:UPPR?
Return value: 1
FUNC:SOURce:STEP:IR:LOWR
To set /inquiry about the LOWER resistance for IR test.
--
Syntax:
Command message: FUNC:SOUR:STEP <sn>:IR:LOWR<resistance >
Query message: FUNC:SOUR:STEP <sn>:IR:LOWR?
--
Data<resistance>:
Data format: float
Data range: 0.1M
Ω
~
50.0G
Ω
Data accuracy: 0.1 M
Ω
Data unit: M
Ω
--
Example:
Set the lower resistance for IR test in STEP1to 1 M
Ω
.