27 | 8 4
RampJudg
CLOSE or ON
CLOSE
When the RampJudg is set to ON, the DC
withstand voltage testing judges whether
the current test value exceeds the set value
of current high limit when performing the
ramp time.
When the RampJudg is set to Close, the DC
withstand voltage testing does not judge
whether the current test value exceeds the
set value of current high limit when
performing the ramp time.
Ramp Time
0.1
~
999.0s,
0 — CLOSE
CLOSE
Set the time required to rise to the set
voltage.
When Ramp Time is set to Close, the
voltage rise is completed during the first
cycle of the test.
DwellTime
0.1
~
999.0s,
0 — CLOSE
CLOSE
Set the time needed for dwell (the current
high limit and low limit are not judged
within the dwell time, but should not
exceed the high limit of the set current
grade).
MeasTime
0.3
~
999.0s,
0 — Continue
3.0S
Set the measure time for the DC
withstanding voltage testing.
0 indicates that the test is continued until
STOP key is pressed.
Fall Time
0.1
~
999.0s,
0 —CLOSE
CLOSE
The time it takes to drop from the set
voltage to low voltage.
CLOSE means that the test time is over and
the voltage output is cut off directly. (The
DUT may be charged).
Table 4-4