Setup items
Range
Default
Explanation
Trg. Mode
Manual
External
Bus
Manual
Set the trigger mode of the instrument to start the
test, only accept the trigger signal in the current
trigger mode.
Trg. Dly
0.0
~
99.9S
0.0S
Set the delay time from the receipt of the trigger
signal to the start of the measurement.
Mea. Mode
Normal
Repeat
Continue
Normal
Normal: Only one test is performed according to
the file settings.
Repeat: Perform the test multiple times as set up.
Continue: Uninterrupted continuous cycle test.
Rpt. Ctn.
0
~
999
0
For the above repeated test mode, repeat count
can be set.
Rpt. Int.
0.0
~
99.9S
0.0S
For repeated and continuous tests, set the time
interval between two tests.
AfterFail
Continue
Restart
Stop
Continue
The test fails, that is, after the instrument reports
FAIL, it can be set to continue the next test, or
press [START] to restart, or press [STOP] first,
then press [START] to start the test.
PassHold
0.2
~
99.9S
0.5S
Set the duration of the buzzer sound when in
PASS.
StepHold
0.2
~
99.9S/key 0.2S
Set the time interval between test steps, the key
means to press [START] to continue test after the
test stopped.
HardAGC
ON/OFF
ON
Set the hardware automatic gain compensation
function to ON or OFF.
SoftAGC
ON/OFF
ON
Set the software automatic gain compensation
function to ON or OFF.
AutoRange
ON/OFF
OFF
Set the withstanding voltage automatic shift
function to ON or OFF.
GFI
ON/OFF/FLOAT
ON
Set the electric shock protection function to ON
or OFF.
Table 4-1
4.2.1.1 Trigger mode
The tester can only start the test after receiving the trigger signal. This tester has four kinds of
trigger modes: manual, external (EXT), BUS, automatic.
The tester can only accept the trigger
signal in the current trigger mode and the trigger signal is only valid under the TEST interface.