69 | 8 4
Data unit: V
--
Example:
Set the starting voltage for breakdown test to 1000V.
Command message: FUNC:BREAK:VOLTS 1000
Query message: FUNC:BREAK:VOLTS?
Return value: 1000
FUNC:BREAKdown:VOLTEnd
To set /inquiry about the VOLTAGE END for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:VOLTE<voltage>
Query message: FUNC: :BREAK:VOLTE?
--
Data<voltage>:
Data format: float
Data range: 50~5000
Data accuracy: 1
Data unit: V
--
Example:
Set the ending voltage for breakdown test to 5000V.
Command message: FUNC:BREAK:VOLTE 5000
Query message: FUNC:BREAK:VOLTE?
Return value: 5000
FUNC: BREAKdown:UPPC
To set /inquiry about the HIGH CURRENT for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:UPPC <current>
Query message: FUNC:BREAK:UPPC?
--
Data<current>:
Data format: float
Data range: 0.0001
~
25mA (DC); 0.001
~
100mA (AC)
Data accuracy: 0.0001mA (DC); 0.001mA (AC)
Data unit: mA
--
Example:
Set the upper current for breakdown test to10mA.
Command message: FUNC:BREAK:UPPC 10
Query message: FUNC:BREAK:UPPC?
Return value: 10.000
FUNC: BREAKdown:LOWC
To set /inquiry about the LOW CURRENT for voltage
breakdown test.
--
Syntax:
Command message: FUNC:BREAK:LOWC<current>
Query message: FUNC:BREAK:LOWC?