![Scientific SME1190 Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/scientific/sme1190/sme1190_user-manual_1217340081.webp)
81 | 8 4
Command message: FETCh:AUTO ON
Query message: FETCh:AUTO?
Return value: ON
FETCh?
Output the test result.
Command Syntax: FETCh? After the tester receives this command, the tester will automatically issue
the test results for all steps until the end of the test.
Return Format: test steps: test mode, actual test voltage (kV) ,actual test current (A), test result;
Example: STEP 1:AC,1.000,1.000e-3,PASS; STEP 2:DC,1.500,0.100e-3,PASS;
Test Steps: 1, test mode: AC, actual test voltage 1kV, actual test current 1mA, test result:PASS;
Test Steps: 2, test mode: DC, actual test voltage1.5kV, actual test current 0.1mA, test result:PASS;
Note: The tester will automatically return each test result by default (per step test result).
5.2.8 Other Commands
*IDN
Inquiry about the model and version information of the instrument.
Return message: <manufacturer>,<model>,<firmware><NL^END>
Where,
<manufacturer> Gets the manufacturer name (Tonghui)
<model> Gets the machine model (such as SME1190/9110A)
<firmware> Gets the version number of firmware (such as Version1.0.5)
For example: *IDN?
FUNC:START Start Test
*STOP
Stop Test