Pilot²
- 38 -
Apex locator
(continuation of the table)
Measurement tends to
show apex too early
or the signal is at a
maximum
Subsidiary current
or
high conductivity
Remove moisture from the crown or the
‘cavity bottom’. Are there side canals? If
necessary, rinse with a saline solution.
Wireless foot switch
No function
Batteries empty
Open the battery compartment under
the switch’s base plate and change the
batteries. Do not use rechargeable
batteries.
No function
Switch not
recognized.
Connect the switch with the device / see
Setup
No function
Signal interference
due to strong
electromagnetic
radiation. (EMC)
Switch off other devices (such as mobile
phones) in the vicinity. Check the
environment.
(A cable switch can be used in
particularly exposed areas)
DownPack
The instrument does not
heat up
Instrument
damaged
Insert new heating tips
(no third-party products!)
Instrument turns
Use a tool for firm clamping
BackFill
No gutta-percha at the tip
of the needle
Handpiece too cold Has the waiting time been adhered to?
Is the heating chamber perceptibly
warm?
Pellet is used up
Is the piston fully pressed in? Insert a
new pellet
Lock-in slips
Turn the rotary knob clockwise as far as
possible (see 11d)
Piston is blocked (cannot
be pulled back during
cleaning)
Gutta-percha
residues
The residues stick to the piston, the gun
must be emptied completely while it is
still warm
Ultrasound
Poor power output
Tip not tightened,
deformed or worn
out.
Thread may be
wrong
Tighten the instrument with a wrench or
replace the instrument.
Caution Satelec (Acteon) and EMS
threads are different.
Damage to the handpiece is possible
No function
Tip not tightened,
cable may be
defective
Tighten the instrument with a wrench.
Contact the service team
If the problem cannot be solved, please contact your dealer or Schlumbohm directly for
advice.
If the device switches itself off, this may indicate a malfunction. Please contact the
manufacturer. Avoid mechanical damage. Do not open the device yourself
.