Pilot²
- 16 -
•
Ensure that the file clamp of the apex cable has been put together correctly after
preparation and that the contact has been screwed-in tightly.
•
*Ultrasonic module: Do not use any deformed or worn instruments. At first, always
choose a very low ultrasonic power output and only increase the energy when
necessary. The ultrasonic device is intended for intermittent (interrupted)
operation. To keep heating to a minimum, operation should be limited to 1 minute
at maximum power and to 4 minutes at minimum power.
•
*Ultrasonic module: Please note that the ultrasonic instruments heat up during
operation. Make sure, therefore, that there is appropriate external cooling, if
necessary.
Compatibility
•
Endo files: You can use all US registered available nickel titanium files with a
standard ISO shaft, tools-coupling ISO 14457:2017.
Do not use instrument parameters out of file manufacturer’s specifications.
•
DownPack heating tips: Only use the original tips which are available from the
manufacturer Meta Biomed.
•
Backfill needles: Only use the original needles which are available from the
manufacturer. Obtura, Young Innovations
•
*Ultrasonic module: Ultrasonic tips: You can use all US registered endo
instruments from Satelec Acteon
•
Warning, do not use non-FDA-registered accessories
General information
•
Keep this user manual and all information safe on the device.
•
Keep the documents for the entire product life cycle.
•
The operator is obliged to report all incidents within the meaning of the current
regulations for medical devices, as well as any information on risks, to the
manufacturer.
2. First steps
2.1.
Assembly
Please first compare the components delivered with the enclosed shipping documents and
the corresponding serial or LOT numbers. Check that the display glass is undamaged.
Please note that all components are supplied non-sterile and not disinfected
(see
chapter 18). Even the brand-new device needs to be processed before it is used for the first
time.
The following conditions should be considered when installing the device:
•
The support surface must be level and made of non-combustible material.
•
The device must not be installed in damp places. Do not use the device
in areas when liquids have been spilt on the floor.
•
Do not expose the device to direct or indirect heat sources. (E.g. sun or radiator)
•
Only charge or operate the device when it is at room temperature (do not exceed
max. +40°C)!