5. Obturation
5.1
Setting the DownPack temperature
5.2.
Setting the BackFill temperature
The temperatures shown are only examples
.
Here, the remaining run time of the BackFill gun
is displayed.
5.1.
DownPack
In this menu, you can select the temperature of the heating tip. The temperature column is
set to the desired temperature (touch display). Start the heating process by pressing the foot
switch. An acoustic signal sounds when the desired temperature is reached. Releasing the
foot switch (after a few seconds) will stop the heating process (second acoustic signal).
If you press the foot switch for a longer time, the heating process is automatically stopped
for safety reasons. The shut-off time depends on the chosen temperature. At maximum power,
the device stops after 5 seconds. At low power, the device stops after 40 seconds at the
latest.
Do not press several times in quick succession. This will make the needle hotter than desired.
During the heating process, the connected BackFill gun is switched off for a short time.
Please also refer to chapter 5.2.
5.2.
BackFill
You can select the temperature of the gun’s heating in this menu. The BackFill gun must be
connected. The temperature column is set to the desired temperature using the touch display.
The device is heated as soon as a temperature has been chosen. The ‘Heating up’ message
appears. When the temperature is reached, an acoustic signal sounds and the auto-off time
of max.
15 minutes
is displayed. After the time has elapsed, the heating process is
automatically switched off again (acoustic signal). You can switch off the heating process
manually by pressing the
OFF
button at the bottom of the temperature column.
Do not leave the device unattended during the BackFill operation.
During obturation (D-Pack and BackFill), the patient must no longer be connected to the apex
cable!