Pilot²
Doc. No. 610 2221 v07
- 15 -
(*applicable with the ultrasonic extension module)
•
Do not expose the device to direct or indirect heat sources.
•
The device may not be used in an environment with free oxygen, explosive or
inflammable gases or flammable liquids.
•
The EndoPilot² should not be installed near devices emitting electromagnetic
radiation so as not influence the correct length determination. Switch off mobile
phones in the immediate vicinity during treatment. (See chapter 17 EMC)
•
Do not cover the device with cloths or foils. Flammable materials may be damaged
or even ignited if the DownPack function is activated unintentionally.
•
Ensure that the rooms in which the device is used are equipped with smoke
detectors. National fire protection regulations must be adhered to.
•
Never leave the appliance unattended when in use.
•
Ensure that the foot switch cannot be pressed unintentionally, for example by a chair
or trolley.
•
The signal of the wireless foot switch is transmitted in an encrypted form. This
technology ensures a secure connection between the foot switch and the device.
This prevents unintentional operation of one device with the foot switch of another
device. Do not operate mobile phones or devices with strong electromagnetic
radiation in the immediate vicinity of the device. This may impair the wireless foot
switch’s function in individual cases.
•
The device does not contain any life-supporting functions. The continued
application in the event of device failure will likely be impossible. This failure will not
endanger the patient’s life. Make sure that the treatment can also be completed in
the event of device failure.
Device components and accessories
•
The power supply has a safety-relevant function. Only use the supplied, medically
approved, original power supply unit!
•
Follow the file manufacturer’s instructions for use and disposal of the endodontic
files.
•
The accuracy of the length determination, the torque and the speed are only
guaranteed when using the EndoPilot² 1:1 contra-angle.
•
An exact length determination may not always be possible due to abnormal or
unusual canal morphology (blocked or fractured canal).
•
The tolerance for torque and speed is 10%.
•
The DownPack handpiece and BackFill gun become hot. There is a risk of burning,
damage to the environment and fire.
•
Place the DownPack handpiece and the BackFill gun back into the holder
immediately after use.
•
Only place the BackFill gun's thermal protector on the gun immediately before use
in the mouth, as this keeps it cool. Remove the thermal protector from the hot gun
after use.
•
To avoid the leading-in of external voltages, the hand-pieces and the lip clip must
not be put down on electrically conductive surfaces.
•
Always remove the lip clip from the patient's mouth when the apex measurement is
not required. The lip clip must not be in the patient's mouth when using ultrasound,
BackFill or DownPack. Never place the lip clip, file clamp and motor on conductive
surfaces. Always return the motor to the handpiece holder. Always place the lip clip
on the retaining provided for this purpose.