Pilot²
Doc. No. 610 2221 v07
- 17 -
(*applicable with the ultrasonic extension module)
•
The ambient temperature must be within the prescribed limits.
(See chapter 16). Avoid heating up to above 60°C in any case!
•
The device must not be installed near free oxygen, flammable gas mixtures or
liquids (e.g. in operating theatre or emergency areas).
•
The EndoPilot² should not be installed near devices emitting electromagnetic
radiation so as not influence the correct length determination.
•
Place the foot switch in such a way that it will be easy to operate.
•
Make sure that the foot switch cannot be activated unintentionally.
•
Place the device in such a way that the power supply cable can be pulled out of the
device when necessary.
2.2.
Holders for the handpieces
The holders provide a safe position for the applied parts. Insert the retainer for the apex cable
laterally into the hole at right handpiece holder.
You can upgrade the device with additional functions (example: left arm with a holder for
DownPack and BackFill). For the assembly of additional holders as well as for the
arrangement of the handpieces, please follow the assembly instructions provided separately.
*Ultrasonic module: A double holder is mounted on the extension module for the ultrasonic
handpiece and the motor.
2.3.
Connection
All connections are plugged in and must
not
be twisted! Care must be taken to ensure that
the plug’s groove fits into the socket’s groove. The ‘Push and Pull’ connections for the
handpieces are color-coded (the numbers refer to the illustrations on the inside cover page).
Figure
Connection Use
1a
blue
Motor
1b
green
Apex cable, connection to the patient (lip clip)
1c
black
Power supply unit
1d
Slot
microSD card
1e
red
BackFill gun, optional
1f
blue
D-Pack, optional, please do not insert the motor here!
Insert the EndoPilot² microSD card into the SD slot before first use.
(Insert the card carefully, do not use sharp tools).
*Ultrasonic module: The ‘Push and Pull’ connections for the handpieces are color-coded as
for the EndoPilot²:
Figure
Connection Use
16a
grey
Ultrasonic handpiece cable and ultrasonic handpiece
When connecting the ultrasonic handpiece to the ultrasonic cable, ensure that the handpiece
is not twisted during connection. If the EndoPilot² control unit is not already equipped with
the ultrasonic module, it can be done later. The device can be upgraded in the factory. Your
dealer will advise you how to reship the device.