Pilot²
Doc. No. 610 2221 v07
- 35 -
(*applicable with the ultrasonic extension module)
12.2.
Maintenance
You can find detailed information on the processing of individual components in chapter 18
(Cleaning, disinfection and sterilization) of the mentioned preparation instructions.
It is imperative that you follow the following instructions:
- Check the connection cable and plug connections every 6 months.
- If the wireless foot switch is not used for a long time, the batteries must be removed.
- Do not use rechargeable batteries for the foot switch (only batteries of type AAA with 1.5V).
- The EndoPilot² does not contain any components that can be repaired on the spot.
- Modifying or opening the device voids the warranty.
- Repairs may only be carried out by the manufacturer!
- Even for infrequent use, the device should be charged every 6 months.
- Only use the intended power supply unit!
- A battery change is planned after 4 years. Heavily aged batteries can pose a safety risk.
Caution: Only the manufacturer or an authorized partner may change the battery.
Caution!
The motor must not be lubricated or oiled under any circumstances! When
servicing the contra-angle, make sure that no lubricant or cleaning agents enter the
motor! Allow excess oil to drip out of the contra-angle before operation. For this
purpose, set the contra-angle down in a vertical position.
The contra-angle should be oiled with oil spray for contra-angles immediately after
application (before reprocessing) to remove penetrated treatment fluids such as sodium
hypochlorite.
Before operating the ultrasonic handpiece, check for damage to the insulation and check
whether the oscillator’s axis is firmly anchored in the housing before each application. Do not
use any damaged ultrasonic handpieces or cables.
Make sure that the connectors are dry.
12.3.
Transport
Prevent the device from falling. The device contains a lithium-ion battery (Li-Ion battery)
Power output: 48 Wh. A hard fall may lead to mechanical damage to the device and the
battery unit. The inserted battery may cause fires and injuries if handled incorrectly. The
device must not be heated up to above 60°C, burned, immersed in liquids or dismantled.
If possible, please use the manufacturer's packaging or sufficiently strong packaging for
shipping the device. Please follow the applicable shipping regulations.
Switch off the device prior to shipping and remove the microSD card. Enclose these in a
clearly visible way. Pack the unit so that the on/off switch on the rear side of the device cannot
be activated unintentionally during transport.
Ensure that all components are disinfected prior to shipping (see processing instructions in
chapter 18).
Dirty and contaminated products must not be shipped.
For shipment please follow the current transport regulations for equipment with Lithium-Ion-
Batteries (UN3481).