Schambeck SFD GmbH
Refractive Index Detector RI2012
RI2012 V1.0 06-2013.EN / 28.10.2013
60
http://www.schambeck-sfd.com
To find out which value is displayed in which column press the
ENTER
button.
The display will change to the following view:
To go back to the former view press the
ENTER
button again.
Diff
represents the difference voltage (in mV) and corresponds to the
difference of intensity reaching the sample side and the reference side of
the light sensor.
Summ
represents the sum voltage (in mV) and corresponds to the light intensity
reaching both sides of the light sensor.
Smpl
represents the sample voltage which corresponds to the light intensity
reaching the sample side of the light sensor.
Rfrn
represents the reference voltage which corresponds to the light intensity
reaching the reference side of the light sensor.