Schambeck SFD GmbH
Refractive Index Detector RI2012
RI2012 V1.0 06-2013.EN / 28.10.2013
35
http://www.schambeck-sfd.com
15 T
HE
S
ERVICE
M
ODE
Since version 3.x the internal firmware of the RI2012 refractive index detector contains the
service mode. This mode allows the user to check more detailed system parameters for
example in case of a malfunction of the instrument to get more detailed information to call
the technical support.
To access service mode press the keys
ArrowLeft
(
◄
) and
ArrowRight
(
►
) at the same
time for about three seconds. The display will change from normal measuring mode
(NormMode) to service mode (ServMode).
The numerical values for temperature, optical balance and the current detector signal are
now displayed with three decimals. In this mode the status line is used to navigate between
several sub menus.
You might choose between:
NormMode
switch back to normal measuring mode
SwHwInfo
switch to Software/Hardware information
ViewFine
switch to view fine system analysis mode
AdjuDACs
switch to adjust mode for digital analog converter
Calibr’n
switch to calibration mode
To switch back to the normal measuring mode press the
MENU
button until the NormMode
statement is shown in the status line.
Now press the
ENTER
button to leave the service mode.