![Schambeck RI 2012 Скачать руководство пользователя страница 26](http://html1.mh-extra.com/html/schambeck/ri-2012/ri-2012_operation-manual_1210122026.webp)
Schambeck SFD GmbH
Refractive Index Detector RI2012
RI2012 V1.0 06-2013.EN / 28.10.2013
24
http://www.schambeck-sfd.com
14.3 T
HE
N
ORMAL
M
EASURING
M
ODE
(N
ORM
M
ODE
)
After performing the internal test procedure the detector will switch automatically to the
normal measuring mode (NormMode). Now internally several tasks are running:
Data acquisition
Data processing
Data output
Internal instrument control
The display will show the following information:
Line 1
shows the current temperature of the optical bench in [°C].
Line 2
shows the optical balance between the two photo diodes of the light sensor in [%%]
which corresponds with the relative position of the light beam on the surface of the
light sensor.
Line 3
shows the current detector signal or error messages (in case of intensity alarms).
The shown signal corresponds with the signal which can be measured at the
integrator output (IntOut). In the normal measuring mode (NormMode) the signal is
displayed with one decimal.
Line 4
is the status line and shows the current status of different detector parameters.
Several parameters are shown. Every few seconds the display will change that you
will have a complete overview of all parameters in a short time.