![Schambeck RI 2012 Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/schambeck/ri-2012/ri-2012_operation-manual_1210122050.webp)
Schambeck SFD GmbH
Refractive Index Detector RI2012
RI2012 V1.0 06-2013.EN / 28.10.2013
48
http://www.schambeck-sfd.com
16.2 E
RROR
M
ESSAGES
In case of some malfunction the RI2012 refractive index detector may display an error
message. The most error will appear directly after switching on the instrument.
16.2.1 M
ALFUNCTION OF THE
T
EMPERATURE
S
ENSOR
There are two possible error messages indicating problems with the temperature sensor.
A more detailed description of the temperature control of the RI2012 can be found in
chapter 16.8
The message “noTS” (no temperature sensor) will appear for that reason that the
temperature sensor is not connected to the circuit board or not working.
The message “!” is shown if the temperature of the optical bench is outside the range
between 9 °C and 64 °C. Beside the displayed message an acoustic signal will occur.