Schambeck SFD GmbH
Refractive Index Detector RI2012
RI2012 V1.0 06-2013.EN / 28.10.2013
41
http://www.schambeck-sfd.com
15.4.1
H
OW TO
CALIBRATE
…
To activate the calibration mode press the
MENU
button until the statement “Calibr’n” is
shown in the status line.
By pressing the
ENTER
button the calibration mode is opened. The following information
will be shown in the display. DifW = Difference with water, SumW = Summ with water,
Dift = Difference with test solution, SumW = Summ with test solution,
By pressing the
ENTER
button you can see the current SPAN.
To start the calibration press the
ENTER
button. The following information will be shown in
the display.
To start the first step of calibration press the
ENTER
button. For the first step of calibration
the valve will be switched automatically into purge mode (the red purge
LED
will light up).
The display will show the line “nowWATERwash”.