ONVERTER
This error messages only occur directly after switching on the RI2012 refractive index
detector.
If this message
“noAD” shows up the analog digital converter did not connect to the micro
controller. Switch of the instrument, wait for ten seconds and try to start again. If the
problem remains contact the technical support.
After switching to service mode you will also notice the error message “noTS”. As the micro
controller also controls the temperature of the optical bench this error message shows up
but does not mean that there is a problem with the temperature sensor.