SBS System – SB-1000 Operation
Error
Code
Message
Description
K
Checked at the end of Automatic mode balance
cycle.
Cleared manually.
Abnormal Auto Balance cycle.
Occurs when a balance cycle finishes, but an
error occurred and was cleared during the cycle.
L
Checked continuously.
Clears automatically.
Unable to measure vibration.
Control may need repair.
M
Checked at power-up.
Does not clear.
Controller has older logic.
Recommend factory update.
N
Checked at power-up.
Does not clear.
Controller has logic problem.
Recommend factory repair.
Factory defaults
Holding the
button down during power-up resets all configurations back to factory defaults. To confirm
the default action, the display will show the
screen icon until the button is released. This action is not
allowed if the FPI input is active on the CNC hardwire interface.
Factory default settings are as follows:
Limit (0.40)
Tolerance (1.20)
Critical (20.00)
Critical RPM (OFF)
Manual RPM (500)
Mode (Auto Balance)
Scale Direction (Same)
Manual Balance Type (2 weight)
Fixed Positions (4)
Test Weight (0.1)
Weight Units (g)
Additive/Absolute Weight Display Mode (+)