SBS System – SB-1000 Operation
13
Pressing the
Setup button will save data and advance to the next setting in the menu. From the last
setting screen in the menu, the unit will exit setup mode and return to the Main screen. To advance to the
next setting without changing a setting, just press the
Setup button without using any arrow buttons.
Pressing the Manual button will save data and exit setup mode, returning to the Main screen.
Pressing
Cancel will discard unsaved changes and revert to the previously saved data. If there is no
unsaved data,
Cancel will exit setup mode and return to the Main screen.
A display option is available for FPI mode. This setting controls if the current vibration reading is displayed
or hidden (***.*) when FPI mode is activated.
Automatic Balance Mode
Setup Automatic Mode
There are four settings under the
Setup button in this mode. These settings are described in the
“Preparing to Set Operating Parameters” section. Each press of the
Setup button will present each of
these settings in the following sequence.
The Limit setting is displayed at the first press of the
Setup
button. The Limit indicator flashes to emphasize this mode.
The edit range is 0.02 to (Tolerance – 0.2).
The Tolerance setting is displayed at the next press of the
Setup button. The Tolerance indicator flashes to emphasize
this mode. The edit range is (Limit + 0.2) to (Critical – 0.2).
The
Critical Vibration setting is displayed at the next press of
the
Setup button. The
Critical indicator flashes to
emphasize this mode. The edit range is (Tol 0.2) to 99.99.
The
Critical RPM setting is displayed at the next press of the
Setup button. The
Critical RPM indicator flashes to
emphasize this mode. The edit range is 300 to 30100, and OFF
displays 0.