PR5230 Instrument Manual
Installing the Instrument and Plug-in Cards
Sartorius
EN-73
4.4
Connecting Load Cells
Load cells or analog platforms (e.g. from the CAPP series) can be connected.
The supply voltage is protected from short circuit and overload.
Caution!
Do not shorten the load cell cable. Connect the prepared cable end and roll up the
remaining cable.
Terminal block Connection Description
+ M
+ Meas. + Meas.
voltage/load
cell
output
- M
- Meas. - Meas.
voltage/load
cell
output
+ V
+ Supply
voltage
+ S
+ Sense + Sense
- S
- Sense - Sense
- V
- Supply - Supply
voltage
Ground Ground
4.4.1
Connecting a Load Cell with 4-Wire Cable
Caution!
The cable colors shown above are applicable to the 4-wire PR62.. series load cells.
Before connecting, check the assignment of cable colors in the load cell manual.
Provide links:
between – Supply (-V) and – Sense (-S)
b Supply (+V) and + Sense (+S)
4.4.2
Connecting PR6221 Load Cells
See installation manuals of PR6221 and PR6021/08, -/68.