Commissioning
PR5230 Instrument Manual
EN-116
Sartorius
5.6.3
Operating parameters
Define the basic operating parameters under
-[
Operating parameter
].
Press
to select the specific parameter.
Press
to select the respective values.
Setup/
Operating parameter
Address A
Select A - Z
PIN
******
Enter the access code
Sequencenumber 28
Counter for print jobs
SetTareKey
tare
&
reset tare
tare
&
reset tare, tare
&
tare again, disabled
SetZeroKey
only when not tared
Only when not tared, reset tare on zeroset,
disabled
[Address]
Device address; e.g. for print-out.
[PIN]
The access code can be used to protect the [Setup] from unauthorized operation. Enter a number with up to
6 digits. As long as you are in this menu, the value can be overwritten as required.
SUPER-PIN
If the PIN-Code is lost, the Setup can be unlocked with Super-PIN '212223'.
[Sequencenumber]
The number (counter for print jobs) may appear (selectable) on the print-out too, is incremented automatically
(max. 999999) and can be set here.
[SetTareKey]
The function of the tare key on the keypad can be configured:
[tare
&
reset tare]: the scale is tared if it was not tared previously; otherwise, tare is reset.
[tare
&
tare again]: when you press the [Tare] key, the current value is stored in the tare memory and a net
weight of 0 is displayed. [disabled]: The key has no function.
[SetZeroKey]
[only when not tared]: The function of the zero-setting key on the keypad can be limited to gross mode. [reset
tare on zeroset]: The zero-setting key switches the scale to gross mode automatically. If the zero-setting key
with these settings has no effect, the configured zero-setting range (around the zero-point set with the dead
load) is already utilized due to a previous zero-setting operation and/or automatic zero setting. [disabled]: The
key has no function.
Press
to return to the setup menu. This message is displayed:
Press [Yes] to save the data, or [No] to close the menu without changing data.