Commissioning
PR5230 Instrument Manual
EN-132
Sartorius
5.9.2
Setup
This event log memory contains events, which are stored as altered data in EEPROM.
Code Events
1
Wrong PIN has been entered.
2
EEPROM has been deleted.
10
Start new adjustment.
11
Set dead load with mV/V.
12
Set span with mV/V.
13
New dead load with weight.
14
New SPAN with weight.
15
New full scale.
16
New scale interval.
17
Memory for adjustment.
18 Changes
undone.
100
Serial parameters stored.
101
Change of serial interface assignment.
200
Serial assign stored.
300
Digital input/output changed.
400
Parameter limits changed.
500
Analog out parameters changed.
600
Field bus parameters changed.
700
Network parameters changed.
800
Software parameters changed.
900
Printer parameters changed.
1000
ADC parameters changed after download via HTTP.
1001
Core EEPROM changed after download via HTTP.
5.9.3
Indicator
Code Events
1
Internal arithmetic error
2 Overload
3
No valid ADC values
6
No sense input voltage
7
Negative input voltage (wrong polarity)
9
Faulty communication with ADC
5.9.4
Powerfail
Only the instrument switch-on is recorded.
5.10
Backup of EAROM
This function will be selected in the main menu under [Backup of EAROM].
The configuration data of the two EAROMs can be saved for back-up on the PC and downloaded, if necessary.
The content of the two EAROMs in the system can be saved for back-up on the PC and downloaded, if
necessary.