Fieldbus Interface
PR5230 Instrument Manual
EN-166
Sartorius
9.1.1
Write Window (Input Area)
Data transmission from the master to the slave (PR5230) is in this window.
Byte 0 Write data: MSB
Byte 1
"
Byte 2
"
Byte 3 Write data: LSB
Byte 4 Read_Value _Select
Byte 5 Write_Value _Select
Byte 6 Direct control bits
The first four bytes are used only for writing a data value.
The register number is in byte 5.
Bytes 6 and 7 contain bits in direct access independent of the
write data.
The command is executed after a 0-1 transition of the
corresponding bit.
Byte 7 Direct control bits
9.1.2
Read Window (Output Area)
Data transmission from the slave (PR5230) to the master is in this window.
Byte 0 Read data: MSB
Byte 1
"
Byte 2
"
Byte 3
Read data: LSB
Byte 4
Read_Value _Selected
Byte 5
General system bits:
- Write_Active
- power_fail
- analog error..
.
Byte 6
Status bits
The first four bytes are used for reading a data value.
The register number of data is mirrored in byte 4 by the write
window, when data is available.
Bytes 5, 6 and 7 contain status bits independent of the read
data.
Byte 7
Status bits
9.1.3
Reading and Writing Data
The number of data exceeds the size of the read/write window by far. For this reason, data is addressed with
Write_Value_Select
a
nd
Read_Value_Select.
To do this, the first six bytes of the write window and the first
five bytes of the read window are required. These can be used by the master to write data in PR5410: e.g., a
limit value is set to 100kg. The master can also read out weight values or other data from the PR5410. For this
purpose, the write and the read window are always required. Safe data exchange is ensured by a write and a
read procedure.
For reading status bits, and for writing direct control bits, however, no procedure is required. The general
system bits and the status bits are always provided and need not be requested. The direct control bits are also
available continuously.
Procedure for reading data:
1.
Writing the register number as
Read_Value _Select
in b
yte 4
of the
write window
(e.g. net weight).
2.
Waiting, until the
Read_Value _Selected
in b
yte 4
of the
read window
is equal to the
Read_Value _Select
in b
yte 4
of the
write window
.
3.
Now the value is available in bytes 0 to 3.
Procedure for writing data:
1.
Waiting, until
Write_Active = 0
in the
read window
(ready to receive new data).
2.
Writing the value in
bytes 0 to 3
.
3.
Writing the register number in
b
yte 5 (Write_Value_Select)
.
4.
Waiting, until
Write_Active = 1
(acknowledges data reception).
5.
Writing 0 in b
yte 5
(Write_Value_Select)
->
Write_Active
goes to 0.