Installing the Instrument and Plug-in Cards
PR5230 Instrument Manual
EN-70
Sartorius
4.3.9
PR1721/47 EtherNet-IP Interface
The EtherNet/IP interface PR1721/47 is a plug-in card for mounting in slot 2, see Chapter 4.3.1. The card is
fitted with a standard RJ
-
45 socket for Ethernet connection and contains a powerful TCP/IP and EtherNet/IP
connecting circuitry with transfer rates of 10 and 100 Mbits/s.
Internal connection
34- pin connector on flat cable for
slot 2
External connection
RJ-45 connecting socket
Transfer rate
10 Mbit/sec and 100 Mbit/sec
Autodetection (10/100, HalfDX/FullDX)
Connection mode
Network
Protocol
EtherNet-IP
Configuration
EDS file (PR1721/47 specific)
Cable
Twisted pairs, screened
e.g. patch cable CAT5
Autolink (straight or crossover)
Cable impedance
150 ¥
Cable length to HUB
Max. 115 m
Potential isolation
Yes
Dimens. (LxWxH)
87 x 55 x 15mm
Weight
125 g
Certificate
EtherNet-IP Specification
ODVA File No. 10286
Test Date: 06.09.2005
Vendor ID 90
See also: www.odva.org
Tested according to: CE, UL & cUL
Caution!
Take care, that switches 1 - 8 are set to OFF.
Note:
The IP-Address and the Subnet-Mask are set at Fieldbus parameter
(see also Chapter 5.6.5 and 9).
The EDS file is stored on the CD (directory ‘Fieldbus’ of the according instrument) supplied with
the unit. The current file is also available for download via the Internet:
http://www.sartorius-mechatronics.com [Downloads].
4.3.9.1
Status-LEDs on Fieldbus Card
LED Meaning
Frequency Color
Flashing
1 Hz green
Fieldbus is initialized and ready for operation.
Flashing
2 Hz green
Fieldbus is not initialized.
Flashing
1 Hz red
Check error in ASIC and FLASH ROM, fieldbus is defect.