PR5230 Instrument Manual
SMA Protocol
Sartorius
EN-159
7.4.5
Reply with ‘A’ and ‘B’ Command
Reply format with ‘A' and 'B' commands (variable length):
<LF><xxx>:<yyyyyy><CR>
The reply format and meaning are:
<LF>
Start of reply from 'A'/'B' command
<xxx>
The field name is three characters long, left-adjusted and filled up with spaces on the
right, if necessary.
The following fields are sent:
“SMA”
level/revision
(reply from ‘A’ command)
“MFG” manufacturer marking
(reply from the 1st ‘B’ command)
“MOD” product / model identification
(reply from the 2nd ‘B’ command)
“REV” software version
(reply from the 3rd ‘B’ command)
“SN_“ serial number
(reply from the 4th ‘B’ command)
“END”: This is always the last inscription field
(reply from the last ‘B’ command)
‘:’
Separator between field name and field content.
<yyyyyy>
The data field contains up to 25 characters.
The SMA field contains <level/revision>
With the following meaning: level = (1, 2, etc.); revision = (1.0; 1.1; etc.)
<CR>
End of reply from 'A'/'B' command
Example:
Command Reply
<LF> A <CR>
<LF>SMA:1/1.0 <CR>
<LF> B <CR>
<LF>MFG:Sartorius <CR>
<LF> B <CR>
<LF>MOD:PR5230 <CR>
<LF> B <CR>
<LF>REV:01.01.9 <CR>
<LF> B <CR>
<LF>SN_:148388723 <CR>
<LF> B <CR>
<LF>END: <CR>
Note:
If the controlling computer sends another 'B' command:
<LF> B <CR> <LF> ? <CR>