![Sartorius BP 110-0CE Скачать руководство пользователя страница 75](http://html1.mh-extra.com/html/sartorius/bp-110-0ce/bp-110-0ce_installation-and-operating-instructions-manual_1199652075.webp)
Model
BP 16000 S
BP 12000 S
BP 34000 P
BP 34
BP 8
Weighing range structure
SuperRange
SuperRange
PolyRange
Weighing capacity
kg
16
12
8/16/34
34
8.1
Readability
g
0.1
0.1
0.1/0.2/0.5 1.0
1.0
Tare range (by subtraction)
kg
–16
–12
–34
–34
–8.1
Reproducibility
g
≤
±0.1
≤
±0.1
≤
±0.1/0.2/
≤
±0.5
≤
±0.5
(standard deviation)*
≤
±
0.5
Linearity
g
≤
±0.2
≤
±0.2
≤
±0.3/0.3/
≤
±0.5
≤
±1
0.3
Response time (average)
s
≤
1.5
≤
1.5
≤
1.5
≤
1.5
≤
1
Allowable ambient
operating temperature
°C
0…+40
Operating temperature range
°C
+10…+30
Sensitivity drift
10 °C ... +30 °C
/°C
≤
±2 ·10
–6
≤
±2 ·10
–6
≤
±2 ·10
–6
≤
±2·10
–6
≤
±10 ·10
–6
Pan size
mm
417 x 307
417 x 307
417 x 307
417 x 307 210 x180
Dimensions (WxDxH)
mm
307x538x121
Net weight, approx.
kg
15
15
15
15
3,0
External calibration weight
(of at least accuracy class...)
g
10,000 (F1)
10,000 (F1)
10,000 (F2)
10,000 (F2) 5000 (F2)
Power requirements
V~
Via AC adapter: STNG 6, 230 VAC or 115 VAC, –20% ... +15%
(IP 20 protection)
Frequency
Hz
48–63
Power consumption (average)
VA
8 maximum; 10 average
Hours of operation with fully
charged YRB 06 Z
external battery pack, approx.
h
28
28
28
28
25
Built-in interface
RS-232C-S/V24-V28; 7-bit; parity: even, mark, odd, space;
transmission rates: 150...19,200 baud;
1or 2 stop bits; software/hardware handshake
Standard features/equipment supplied:
AC adapter, varies acc. to country
x
x
x
x
x
Level indicator
x
x
x
x
x
* = standard deviation of the reproducibility acc. to DIN 8120, Part 3
75