Operating the Balance
Warmup Time
To deliver exact results, the balance must warm up for at least 30 minutes after
connection to AC power or after a power outage. Only after this time will the
balance have reached the required operating temperature.
Turning the Display On and Off (Standby Mode)
Press the
e
key
(11)
to turn the display on and off.
Self-Test
After the balance has been turned on, an automatic
self-test of the balance’s electronic circuitry is
performed. At the end of the self-test, a zero readout
is displayed. This means that the balance is ready
to operate.
The display shows the following special codes for your information:
O
displayed in the upper right corner stands for
OFF
The balance was disconnected from AC power
(balance reconnected to AC power or power outage
longer than 3 seconds).
O
displayed in the lower left corner means
standby
The display has been turned off by the
e
key
(11)
.
The balance is now in the ready-to-operate mode and
does not require warmup.
b
means
busy
Once you have turned on the balance, the
b
symbol
will be displayed until you press a key. During
operation, this symbol indicates that the balance
processor is still busy processing a function and will not
accept another command to perform any other
functions at this time.
* = including the Signatories of the Agreement on the European Economic Area
b
o
o
Important Note Concerning Verified Balances Approved for Use as Legal
Measuring Instruments in the EU*:
The last digit on the display is bordered when the verification scale interval “e”
is greater than the scale interval “d.”
20
▼