![Sartorius BP 110-0CE Скачать руководство пользователя страница 46](http://html1.mh-extra.com/html/sartorius/bp-110-0ce/bp-110-0ce_installation-and-operating-instructions-manual_1199652046.webp)
Practical Example: Counting Small Parts
Menu code settings used in the example:
Function
Code
Counting
2 1 4
Ref. qty. and ref. weight
7 1 2
With data ID code
7 2 2
Application:
Counting bulk quantities of items with the same weight
Step/Key
Readout
Data Output/Printout
Place container on pan
+ 22.65 g
cs
c
,
t
0.00 g
cs
Add 10 counted parts
Press
v
to confirm reference rEF
10 pcs
nRef
+
10 pcs
sample quantity
+ 5.6546 g
cs
wRef
+
5.6546 g
+
10 pcs
Fill container with desired
quantity of parts
(without counting them)
+
500 pcs
p
+
500 pcs
Qnt
+
500 pcs
Press
c
to clear the
weight readout and
the reference value
+ 2827.35 g
cs
p
+ 2827.35 g
cs
N
+
2827.35 g
The data ID codes
mean:
nRef
+
10 pcs
reference sample quantity (piece count)
wRef
+
5.6546 g
reference weight
Qnt
+
500 pcs
calculated piece count
N
+
2827.35 g
net weight
46