![Sartorius BP 110-0CE Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/sartorius/bp-110-0ce/bp-110-0ce_installation-and-operating-instructions-manual_1199652038.webp)
Automatic Output of the Tare Memory Data
If you have selected the application program “tare memory” (2 1 6), you can have
your choice of values output by pressing the
v
key:
Code
Last net value (individual value N1)
7 3 1*
Tare memory data (total T1)
7 3 2
Additional Functions
Menu Access Switch Function on Standard Balances
You can define the function of the menu access switch by setting the code for the
balance operating menu to “accessible.” The balance operating menu will then be
accessible at all times, which means that you can change the menu codes at any
time regardless of the setting of the menu access switch.
Access to the balance operating menu
Code
Accessible
8 1 1
Depends on the setting of the menu access switch
8 1 2*
Menu Access Switch Function on Verified Balances
The balance operating menu can also be changed when the balance is being
used as a legal measuring instrument. Codes that are not permitted for operation
of the balance as a legal measuring instrument are blocked and cannot be
selected. For balances of accuracy class
K
, the menu access switch is adjusted
to lock the “External Calibration” function after verification. To do this, the switch
must be moved to the left. A control seal is affixed to the balance to lock the
menu access switch. Unauthorized attempts to remove this seal will irreversibly
damage it. If you break the seal, the validity of the verification will become
void, and you must have your balance re-verified.
Blocking the Keys
You can block all keys on the balance (except for
e
).
Key functions
Code
Accessible
8 3 1*
Blocked
8 3 2
* = factory setting
38