Center the calibration weight on the weighing pan.
The balance then calibrates automatically.
At the end of calibration, the calibration weight
readout and the stability symbol “g” are displayed.
Calibration Test for Balances with a Built-in
Calibration Weight
Menu code selection: 1 9 4
When a zero readout is displayed, press the
q
key
(9)
. The built-in calibration weight is now
internally applied by servomotor. At the same
time,
CAL
is displayed. After the display has
stabilized, the deviation of the current readout from the
target weight (in grams only) is indicated.
If external interference affects the calibration test, you
may obtain a brief display of the error code “Err 02.”
In this case, tare; then press the
q
key again.
Blocking the Calibration/Adjustment Functions
You can block these functions by setting code
1 9 7
(when the menu access switch
(4)
is locked).
Interface Port
Depending on the balance model, unfasten or remove
the protective cap from the data interface port.
– Plug the connector into the interface port
– Secure the connector by tightening the screws
To print or output data, press the
p
key
(7)
.
Important Note
Make absolutely sure to unplug the balance from AC
power before you connect or disconnect a peripheral
device (printer or PC) to or from the interface port.
For information about the data output parameters
and data ID codes, see page 36.
24
▼
▼