![Sartorius BP 110-0CE Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/sartorius/bp-110-0ce/bp-110-0ce_installation-and-operating-instructions-manual_1199652044.webp)
Practical Example: Determination of the Residual Weight in Percent
Menu code settings used in the example:
Function
Code
Weighing in percent
2 1 5
Ref. % and weight
7 1 2
With data ID code
7 2 2
Application:
Quick determination of the dry weight of a sample
Step/Key
Readout
Data Output/Printout
Place container on pan
+
22.65 g
c
,
t
0.00 g
Add prepared sample
+
4.61 g
to container;
v
+
100.0 %
nRef
+
100 %
Wxx%
+
4.61 g
For moisture analysis,
press
t
at this point
0.0 %
Remove container; dry sample
xx.x %
Reweigh sample in container
+
72.5 %
p
+
72.5 %
Prc
+
72.5 %
Press
c
to clear
residual weight readout and
reference value
+
3.34 g
p
+
3.34 g
N
+
3.34 g
The data ID codes
mean:
pRef
+
100 %
reference percentage
Wxx%
+
4.61 g
net reference weight xx%
Prc
+
72.5 %
calculated percentage
N
+
3.34 g
net residual weight
44