Display Modes
You can select the display mode that best meets your individual accuracy
requirements (last numeral). The display increments possible are as follows:
1, 2, 5, 10, 20, 50, etc.
Starting with the basic increments of a weight unit, the display accuracy can
be reduced by as many as three levels so that you will obtain a faster readout with
a reduced display accuracy. The display accuracy is reduced proportionally to
the selected basic increment of a weight unit. To make this concept easier
to understand, the three levels are designated as “rounding factors” in the tables
summarizing the various menu code settings.
Code
Display mode
1st level
2nd level R1
Highest possible accuracy
1 8 1*
3 2 1*
Last numeral blanked when load changes
1
)
1 8 2
3 2 2
Rounding factor 2**
1 8 3
3 2 3
Rounding factor 5**
1 8 4
3 2 4
Rounding factor 10** (last numeral off)
1 8 5
3 2 5
*
*
= factory setting
** = not applicable to verified balances approved for use as legal measuring
instruments in the EU
1
) = applicable only to verified balances approved for use as legal measuring
instruments in the EU