5.5 F2 - noise filter setting
5 Configuration Mode
S122
User /
Service Instructions
5.5
F2 - noise filter setting
The range of stability is from 01 to 15. Enter a setting
appropriate to the environment.
●
The higher the value the quicker the machine reaches
stability but weighing is less accurate.
●
If the environment is unstable enter the value 15.
●
If the environment is stable enter the value 05. This is the
default value.
●
Ensure the hold function (F8) is set to hold 0.
To change the value:
twice
F2
ENTER
fiL 05
fiL 06
ENTER
F2