Error messages
2 Introduction
S122
User /
Service Instructions
12
2.
3
Error messages
E1
When the
scale
is switched on, the weight on the
platform is above or greater than the automatic zero
range (greater than 10% of the capacity of the scale).
Make sure that there is no weight on the platform. If the
error persists, re-calibrate the machine.
E2
When the
scale
is switched on, the weight on the
platform is below or lower than the automatic zero
range (greater than 10% of the capacity of the scale).
Correct service set up procedure has not been
followed. Ensure calibration weight is applied and
removed exactly in accordance with instructions.
E3
Not used.
E4
The weight on the platform is unsteady.
E5
There is a problem with the load cell. Check the wiring
(see page
60
). If the error persists, replace the load cell
(see page
33
).
E6
The internal zero value is greater than 360,000. The
error may occur in calibration mode.
E7
The internal zero value is less than 240,000. The error
may occur in calibration mode.
E8
This message will only appear for
legal for trade
scales
.
If you are configuring the capacity of the
scale
, the
number of divisions cannot exceed 10,000.
The weight is greater than the capacity of the
scale
.
Temporary error. Remove any weight from the platform,
switch off and then on again. If this error persists,
contact your authori
z
ed service agent.
--OL--
------
Содержание S122
Страница 1: ...76104 180 Issue 0 76104 180 21 07 2004 ENGLISH S122 Light Industrial Scale User Service Instructions...
Страница 26: ...Creating a preset tare Example 1lb Reviewing a pre set tare value...
Страница 27: ...Canceling a pre set tare...
Страница 59: ...S122 User Service Instructions 6 2 Wiring Diagrams 6 2 Main board connections 6 2 Main board connections...
Страница 60: ...6 3 Load cell connections 6 3 Wiring Diagrams S122 User Service Instructions 6 3 Load cell connections...